Installation Guide | 3-25
Alarm Contact Connections
This section describes how to make wiring connections for any alarm de-
vices you want to connect to the control unit.
Standard Alarm Connections
The WDG-V sensor provides five sets of normally open alarm contacts as
follows:
Terminal Block ID
Description
SVC
- Service
Alarm
DV
- Data
Valid
Alarm
ALM1
-
Process Alarm 1 (Configurable)
ALM2
-
Process Alarm 2 (Configurable)
ALM3
-
Process Alarm 3 (Configurable)
The process alarms are referenced as Relay 3, Relay 4 and Relay 5 in the
AMEVision.
Relay 3
-
ALM 1
Relay 4
-
ALM 2
Relay 5
-
ALM 3
Service Alarm
- The service alarm indicates that a critical alarm exists and
the analyzer needs repair. Readings are not valid when this alarm is ac-
tive. The relay contacts are normally-open and are OPEN (de-energized)
when a critical alarm exists.
Data Valid
- The data valid alarm indicates if the concentration readings
are representative of the process. When this alarm is active, concentra-
tion readings may not be representative of the process and should not be
used for control/safety. Examples of this alarm include: Warm-up mode,
in calibration, and diagnostics mode. This alarm differs from the service
alarm in that there is not necessarily a problem with the analyzer when
this alarm is active. The relay contacts are normally-open and are OPEN
(de-energized) when the concentration readings are not valid.
Содержание Thermox WDG-V
Страница 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Страница 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Страница 73: ...AMEVision Display User Interface or Communications 4 5...
Страница 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Страница 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Страница 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Страница 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Страница 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Страница 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Страница 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Страница 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Страница 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Страница 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...