User Manual | 45
ERROR MESSAGES
Error Messages
Error Messages are viewed and acknowledged on the System Settings
screen. When the Status LED on the front panel is red this indicates that
there are system errors or the status of the analyzer has changed. The
System Status page displays error messages that the system has encoun-
tered. See the Run section for more information on the System Settings
page (35).
Chassis Controller Error
Internal electronics failure - Call manufacturer or qualified service repre-
sentative.
Detector Controller Error
Internal electronics failure - Call manufacturer or qualified service repre-
sentative.
Temperature Out of Range
Either the temperature zone is 2 degrees over the set point, or the tem-
perature zone has been disabled. Perform the Warmup procedure to clear
this error.
Carrier Pressure Out of Range
External carrier pressure is not within specified limits. Verify external car-
rier pressure supply.
FID Flame Out Error
The FID flame has gone out. Hydrogen or air mixture may be incorrectly
set. Check support gases then perform the Warmup procedure.
FID Flame Ignition Error
Warmup process could not ignite FID flame. Check support gases then try
Warmup process again.