User Manual | 43
Shut Down
System shut down for maintenance or for any other reason is very sim-
ple. If shut down is due to an emergency or for any safety reason, simply
switch the instrument off using the A/C DIN Switch located on the rear
panel of the analyzer (0 depressed is off).
Carrier gas to the instrument must not be stopped until all tem-
peratures are below 50
0
C.
To shut down the analyzer:
From the main menu press the Idle key to stop analysis.
Shut off any support gasses except for the carrier gas.
Press the Cool key and allow the analyzer to cool down for two to three
hours.
Switch the analyzer and all other components off.
Shut off the carrier gas.
!
CAUTION