User Manual | 37
Temperatures and Status Page
Press the Page key to display the Temperatures and Status page. This page
shows status information about the critical heated zones and status of the
detector. No user input is allowed on this page in normal user mode.
Image of Temperatures and Status Screen
CalibrationScreen
Press the Page key to display the calibration information page.
The calibration page is used to calibrate the instrument and shows
information about the most recent run. Name is the gas species to be
measured. Conc. is the concentration (or blend) of the calibration gas
connected to the instrument. Area is the peak area of the impurity
detected during the last analysis run for each impurity type.
Rfactor is the response factor calculated by the system on the most
recent calibration. Rfactor is used to calculate the concentration of
impurities. RT is retention time of the impurity. This is the time corre-
sponding to the top of each peak calculated from the start of the most
recent analysis run.
Idle Mode
Temperatures & Status
FID Meth.
Meas
= 295 #
FID Column = 50
#
Set
50
FID Bias
FID Flame
FID Vout = 49755
= 165 #
------------------------- Pressure -------------------------
Carrier = 23.0