40 | Model ta3000F Gas Purity Monitor
System Settings
V. 2.70
Idle Mode
12:34:25
Edit Print Page
Clock: 10-Jan-03 12:34:25
Print Chromatograms:
+
*
-------------------- System Errors --------------------
Super User Temperature & Status Page
Depressing the “Zero” key will manually zero the detector. Depressing the
“Bias” key will toggle the bias on and off. Depressing the “H2-Prg” key is
disabled.
Depressing the “Edit” key will provide a new menu of options above the
keys to:
1. Toggle the heaters on and off
2. Change the heater set points
Depressing the “Ignite” key will ignite FID flame.
Depressing the “Ignite” key will ignite FID flame.
Super User Calibration Page
The calibration screen has the same appearance as in Normal User mode,
but you now have the option to modify the Rfactor and Concentration
Factor in addition to the Concentration (Conc). For this instrument the
Concentration Factor must be 1 for proper operation.
Concentration and Response Factor
Calculation RF = AREA / CONC
GAS_CONC = AREA / RF
RF
= Response Factor
CONC = Concenration (or blend) of Calibration Gas Cylinder in ppb
Содержание ta3000F