User Manual | 13
Specifications
Model designation and detection limit specifications
Sample Gas Model
Lower Detection Limit
CO2 ppb CH4ppb NMHCppb
Nitrogen
ta3000F-N2
10 10
25
Argon ta3000F-Ar
10 10
25
Helium
ta3000F-He
10 10
25
Oxygen
ta3000F-O2
10 10
25
Hydrogen
ta3000F-H2
10 10
25
Air
ta3000F
10 10
25
Performance
Accuracy
Greater of +1 x LDL or + 10% of readingg
Range
0 to 5 ppm (Higher are available as an option)
Response Time
Length of Analysis is dependant on the application.
Response time is independent of sample concentration.
Ambient Operating Temperature.
60 to 90° F (16 to 32° C)
Sample Compatibility
Specific models available for various applications
Resolution, Display
0.1 ppb
Содержание ta3000F
Страница 47: ...User Manual 47 DIMENSIONAL DRAWINGS Outer Dimensions 3D View ...
Страница 48: ...48 Model ta3000F Gas Purity Monitor ta3ooof Side View 22 115 6 92 ...
Страница 49: ...User Manual 49 24 070 16 70 ta3oooF Top View ...
Страница 50: ...50 Model ta3000F Gas Purity Monitor This Page intentionally left blank ...