Guidelines for prevention of damage
PLEASE STUDY THIS MANUAL THOROUGHLY BEFORE IN-
STALLING OR OPERATING THE INSTRUMENT.
Carrier gas
Detectors can be permanently damaged if operated at temperature with-
out flow. The carrier gas should remain flowing until all heated zones are
below 50o C (2 to 3 hours).
Line voltage
Do not connect the power cord to line voltage without checking the cor-
rect operating voltage on the label located on the rear of the instrument.
Carrier gas purifiers
External purifiers must be completely conditioned with gas flowing
through them before connection to the analyzer. Failure to properly con-
dition purifiers can result in serious contamination.
FID Flameignition
Do not ignite the FID flame unless the FID is at operation temperature.
Moisture will condense in an un-heated FID and will short-out the elec-
trometer input leads.
Shut down
Be sure to turn off all heated zones and shut off the hydrogen supply
to the instrument prior to disconnecting gas supply lines. The carrier
gas should remain flowing until all heated zones are below 50o C (2 to 3
hours). After shutting off carrier gas, immedi- ately cap all open rear panel
gas ports.
Electrical signal cables
Turn power off before connecting or disconnecting electrical signal cables.
Damage to circuit boards could result.
!
CAUTION
Содержание ta3000F
Страница 47: ...User Manual 47 DIMENSIONAL DRAWINGS Outer Dimensions 3D View ...
Страница 48: ...48 Model ta3000F Gas Purity Monitor ta3ooof Side View 22 115 6 92 ...
Страница 49: ...User Manual 49 24 070 16 70 ta3oooF Top View ...
Страница 50: ...50 Model ta3000F Gas Purity Monitor This Page intentionally left blank ...