User Manual | 15
Inlet pressure stability
+ 2%, UHP regulator
Flow
rate
30
cc/min
Purity
Hydrocarbons,
CO,
C
2
O , Moisture < 1 ppm
FID Air
Inlet Fitting
1/16-inch VICI compression fitting
Inlet pressure range
20 psig (1.4 bar)
Inlet pressure stability
+ 2%, regulator required
Flow
rate
230
cc/min
Purity
Total hydrocarbons < 1 ppm, dew
point
<
-
65°
C
Calibration Gas
Inlet Fitting
1/16-inch VICI compression fitting
(Uses
Sample
Inlet)
Cylinder concentration
H
2
, CO, balance nitrogen, depends
on level of detection
Cylinder volume
100 cubic feet minimum (2800L)
Chassis
Dimensions
7” (18cm) H x 16.8” (43cm) W x 26.5”
(67cm)
D
Weight
35
lb.
(15.9
kg)
Power
100 - 120 VAC, 50/60 Hz,; 200 - 240
VAC,
50/60
Hz
Outputs
Display
LCD graphics, backlit, 100mm x 150mm
Printer Port
Concentration, chromatogram, and diagnos
tic
reports
Serial PLC Port
Concentration data and general alarm via
RS232
4 to 20ma (optional)
Concentration signals