14 | Model ta3000F Gas Purity Monitor
Resolution, Communication Ports
0.01 ppb
Carrier Gas Supplies (On-Line Installation)
Inlet Pressure Range
70 to 90 psig (4.8 to 6.1 bar)
Inlet Pressure Stability
± 2%, regulator required
Return Pressure
Atmospheric vent is optimal, ± 0.5 psig
maximum
Flow Rate
50 cc/min minimum, bypass at 50
cc/min.
Temperature
60 to 100 °F (16 to 38 °C), optimum when
maintained
±
2°C
Maximum Impurity Levels Varries by application. An external purifier
may
be
required
Gas Ports
Sample Inlet
1/16-inch VICI compression fitting
Carrier
1/16-inch VICI compression fitting
Sample Vent
1/16-inch VICI compression fitting
FID H
2
1/16-inch
VICI
compression
fitting
FID Air
1/16-inch VICI compression fitting
Aux
1/16-inch VICI compression fitting
FID SHUT-OFF
1/8-inch VICI compression fitting
Sample Gas
Inlet
Fitting
1/16-inch
VICI
compression
fitting
Flow Rate
20 - 100 sccm minimum
Inlet pressure stability
+ 2%, UHP regulator required
Vent pressure
Atmospheric pressure vent is optimal, + 0.5
psig
maximum
Fid Hydrogen
Inlet Fitting
1/16-inch VICI compression fitting
Inlet Pressure Range
30 psig (2.0 bar)