Manual No. M08-372-F1116
User defined test window
1) Test mode: Choose the type of test you wish to create.
2) Test Number.: Select from the list the number of the memory slot in the CT3 where you wish
to load the test.
3) Name: Enter the name of test as you wish it to appear on the screen of the CT3.
Test Parameters window
Enter the desired test parameters. The specific test parameters available will vary depending upon
the test mode chosen.
Now that you have created the test, you may:
1) Save test if desired by clicking on diskette icon button.
2) Click Send Test button to download the test to the CT3.
3) If you decide to change a parameter in a test that has been downloaded to the CT3, you can
make the change and download the test to the same memory slot in the CT3. This will overwrite
the initial program.
Target Hold is added to the user defined downloadable tests. The com spec “xx” field for
“TXTNAME<xx><nnnnnnnn><CR>” is changed from values of 00-04 to 00-05. “05” represents
the added test type Target Hold. The PGM command field “xxxx” is used for the target load value.
The home screen for this test is shown below. Note that the test type is displayed as “TH”.
0:TESTNAME TH nnnn s
TRIGGER:
10.0 g
TARGET LOAD: 500 g
SPEED:
0.5 mm/s
The User Defined target hold test will operate the same as the standard target hold test except that
there will be no ability to modify field values manually. It will essentially operate in a similar
manner to other User Defined tests.