AMETEK Brookfield
Page 33
Manual No. M08-372-F1116
HOLD TIME Full re-start
After a HOLD TIME test is complete, the operator can do a full restart of the test. Pressing the
STOP button at this time will display the setup screen. This places the CT3 back at the HOLD
TIME setup mode. The probe will return to the HOME position. The various setup parameters can
be modified. Pressing START will begin the full test run complete with pre-positioning.
III.9.3 CYCLE COUNT mode pre-positioning
The pre-positioning screen is displayed after the ATTACH PROBE screen. Move the probe to the
desired START position. The displayed value for deformation will be set to 0.0 mm. All deformation
measurements will then be referenced to the trigger point after the probe makes contact with the
sample.
When the probe is retracted upwards between cycles, it will stop at the START position before
descending for the next cycle. At the end of a CYCLE COUNT test, the probe will return to the
START position. The probe will return to the HOME position only if a full restart of the CYCLE
COUNT test is done.
CYCLE COUNT mode Quick re-start
After a CYCLE COUNT test is complete, pressing the START button will re-run the test starting
at the AUTZERO screen.
CYCLE COUNT Full re-start
After a CYCLE COUNT test is complete, the operator can do a full restart of the test. Pressing the
STOP button at this time will display the setup screen. This places the CT3 back at the CYCLE
COUNT setup mode. The probe will return to the HOME position. The various setup parameters
can be modified. Pressing START will begin the full test run complete with pre-positioning.
III.9.4 BLOOM mode pre-positioning
The pre-positioning screen is displayed after the ATTACH PROBE screen. Move the probe to the
desired START position. The displayed value for deformation will be set to 0.0 mm. All deformation
measurements will then be referenced to the position when the probe makes contact with the sample
and the trigger point is reached.
At the end of a BLOOM test, the probe will return to the START position. The probe will return
to the HOME position only if the full restart of the BLOOM test is done.
BLOOM mode Quick re-start
After a BLOOM test is complete, pressing the START button will re-run the test starting at the
AUTZERO screen.
BLOOM Full re-start
After a BLOOM test is complete, the operator can do a full restart of the test. Pressing the STOP
button at this time will display the setup screen. This places the CT3 back at the BLOOM setup
mode. The probe returns to the HOME position. The various setup parameters can be modified.
Pressing START will begin the full test run complete with pre-positioning.
III.9.5 TPA mode pre-positioning
The pre-positioning screen is displayed after the ATTACH PROBE screen. Move the probe to the
desired START position. The displayed value for deformation will be set to 0.0 mm. All deformation
measurements will then be referenced to the position when the probe makes contact with the sample
and the trigger point is reached.
Содержание BROOKFIELD CT3
Страница 2: ......