The following sequence of display screens summarizes the Target Hold setup procedure:
TEST:TRGT HLD: 210.0 s
TRIGGER:
10.0 g
TRGT LOAD: 500 g
SPEED:
0.5 mm/s
* ATTACH PROBE *
PRESS START
TO CONTINUE
* ADJUST POSITION
USING SCROLL SELECT
* PRESS START
DISTANCE
123.5
mm
AUTOZEROING
TEST:TRGT HLD: 210.0 s
TRIGGER:
0.098 N
TRGT LOAD: 4.9 N
SPEED:
0.5 mm/s
Figure III.29
The CT3 displays the following information while the Target Hold test is being run.: TIME AT
TARGET, DEFORMATION, and the current measured LOAD. The sequence of events, and the
corresponding display fields, is as follows.
The display begins with the TIME at 0 while the probe descends at the selected PRE-SPEED (assigned
by the user in the GLOBAL PARAMETERS screen described elsewhere). Pressing and holding
the START button prior to reaching the TRIGGER load will override the PRE-SPEED setting for
the probe with a default speed of 4.5 mm/sec (known as jogging). The DISTANCE remains at 0.0
until the TRIGGER load value is reached when the probe makes contact with the sample. The
LOAD field then starts to display the actual force measured by the CT3.
TEST: TRGT HLD
TIME:
210 s
DISTANCE:
0.0 mm
LOAD:
0.0 g
TEST: TRGT HLD
TIME:
210 s
DISTANCE:
0.0 mm
LOAD:
0.00 N
Figure III.30
When the measured LOAD reaches the assigned TRIGGER value, the probe speed is changed to
the assigned SPEED from the Target Hold setup screen. As the probe approaches the target load
at the pre-assigned speed, there will be a fixed deceleration algorithm that reduces the speed of the
probe until it reaches the target load, based on a percentage of the load cell full scale. Therefore,
the pre-assigned probe speed only represents the maximum allowable speed. For example, if the
applied load, when the TRIGGER point is reached, is sufficiently close to the target hold load, then