Manual No. M08-372-F1116
The operator sets the trigger value, target deformation and the test speed using the Tension menu
screen:
TEST: TENSION
TRIGGER: XXX.X g
DEFORMATION: XXX.X mm
SPEED: XX.X mm/s
Figure III.22
After pressing Start the operator is presented with this reminder screen:
*ATTACH TENSION
FIXTURES
*PRESS START
TO CONTINUE
Figure III.23
Be sure grips or other sample fixtures are secure and aligned then press START. The Scroll/Select
knob now becomes a tool for adjusting the position of the grips so that the sample can be clamped.
Rotate knob to move 1mm/click or hold knob down to descend at 4.5mm/s:
*ADJUST POSITION
USING SCROLL/SELECT
*PRESS START
DEFORMATION:XXX.X mm
Figure III.24
The deformation value shown in this screen is the current distance from the top Home position and
as such shows how much travel is possible once the test begins. Be sure this value is greater than
the sum of the distance necessary to reach the trigger plus the programmed target deformation.
When clamping the sample try to exert only minimal lateral loads on the probe shaft. Pressing Start
as shown below will start the test:
*CLAMP SAMPLE
*PRESS START
TO CONTINUE
Figure III.25
During the test the running screen is seen as shown below:
TEST: TENSION
DEFORMATION:XXXX.X mm
LOAD: XXXX.X g
Figure III.26