AMETEK Brookfield
Page 36
Manual No. M08-372-F1116
of the acceptable range, contact AMETEK Brookfield or an authorized dealer for information on
repair/calibration services (see Appendix D for AMETEK Brookfield locations).
Load Range
Load Accuracy
Model
Grams
CT3-100g
0 – 100g
0.5
CT3-1000g
0 – 1,000g
5.0
CT3-1500g
0 – 1,500g
7.5
CT3-4500g
0 – 4,500g
22.5
CT310kg 0-10kg
50
CT325kg 0-25kg
125g
CT350kg 0-50kg
250g
Table III.2
Example of Calibration Verification
A 4500 gram CT3 Texture Analyzer, according to Table III.2, has an accuracy of ±22.5 grams. The
spreadsheet shown below can be used to easily evaluate a calibration check using the appropriate
Brookfield calibration weight set.
Calibration Spreadsheet
CT3 TEXTURE ANALYZER
serial # 6514047
Load Cell Capacity 4500
#
1
2
3
4
5
Load Cell Accuracy
Mass 999.84 994.04 993.96 994.05 497.12
22.5g
Serial #
59
87
92
106
153
Weight
1 999.84
1 & 2
1993.88
1, 2 & 3
2987.84
1, 2, 3 & 4 3981.89
1 - 5
4479.00
INSTRUCTIONS FOR USE
The CT3 Calibration template can be downloaded from our website,
www.brookfieldengineering.com/products/texture-analysis/cts.asp
Low Limit
High Limit
Reading Difference
977.34 999.5 1022.34 -0.3
1971.38 1993.5 2016.38 -0.4
2965.34 2987.5 3010.34 -0.3
3959.39 3981.5 4004.39 -0.4
4456.51 4479.5 4501.51
0.5
Enter the serial number of your CT3.
Enter the load cell in your CT3. The load cell accuracy is automatically calculated from this
value.
Enter the mass and serial numbers of the weights in the same order that you add them to the
hanger.
Enter the load value shown in the CT3 display as each weight is added.
Calibration is satisfactory if all values in the “Difference” column are smaller than the “Load
Cell Accuracy” value.