5.2. Endoscope specifications
Insertion portion
aScope 4 RhinoLaryngo Intervention
Bending section
1
[°]
130 ,130
Insertion cord diameter [mm, (”)]
5.0 (0.20)
Distal end diameter [mm, (”)]
5.4 (0.21)
Maximum diameter of insertion portion [mm, (”)]
5.5 (0.22)
Minimum tracheostomy tube size (ID) [mm]
6.0
Working length [mm, (”)]
350 (13.8)
Channel
aScope 4 RhinoLaryngo Intervention
Minimum instrument channel width
2
[mm, (”)]
2.0 (0.079)
Storage and transportation
aScope 4 RhinoLaryngo Intervention
Transportation temperature [°C, (°F)]
10 ~ 40 (50 ~ 104)
Recommended storage temperature
3
[°C, (°F)]
10 ~ 25 (50 ~ 77)
Relative humidity [%]
30 ~ 85
Atmospheric pressure [kPa]
80 ~ 109
Optical System
aScope 4 RhinoLaryngo Intervention
Field of View [°]
85
Depth of Field [mm]
6 - 50
Illumination method
LED
Suction connector
aScope 4 RhinoLaryngo Intervention
Connecting tube ID [mm]
Ø7 +/- 1
Sterilisation
aScope 4 RhinoLaryngo Intervention
Method of sterilisation
ETO
Operating environment
aScope 4 RhinoLaryngo Intervention
Temperature [°C, (°F)]
10 ~ 40 (50 ~ 104)
Relative humidity [%]
30 ~ 85
Atmospheric pressure [kPa]
80 ~ 109
Altitude [m]
≤ 2000
1.
Please be aware that the bending angle can be affected if the insertion cord is not kept straight.
2.
There is no guarantee that accessories selected solely using this minimum instrument channel
width will be compatible in combination.
3.
Storage under higher temperatures may impact shelf life.
10