D000161-00 Rev. C
38
APPENDIX B: EMC STATEMENT
Warning:
•
The use of accessories, transducers and cables other than those specified by AlterG,
Incorporated, may result in increased EMISSIONS or decreased IMMUNITY of the
EQUIPMENT.
•
This EQUIPMENT should not be used adjacent to or stacked with other equipment and
that if adjacent or stacked use is necessary, the EQUIPMENT should be observed to
verify normal operation in the configuration in which it will be used.
Guidance and manufacturer’s declaration - electromagnetic emissions
The EQUIPMENT is intended for use in the electromagnetic environment specified below. The
customer or the user of the EQUIPMENT should assure that it is used in such an environment.
Emissions Test
Compliance
RF emissions
CISPR 11
Group 1
The equipment uses RF energy only
for its internal function. Therefore, its
RF emissions are very low and are
not likely to cause any interference
in nearby electronic equipment.
RF emissions CISPR 11
QP-Horizontal
QP-Vertical
Freq-
uency
MHz
Quasi Pk
FS
dB(uV/m)
Limit@
10m
dB(uV/m)
Margin
dB
RA
dB(uV)
CF
dB
AG
dB
AF
dB(l/m)
49.95
31.5
40.0
-8.5
56.5
0.7
32.0
6.3
156.5
23.3
40.0
-16.7
44.5
1.3
32.0
9.5
30.8
28.4
40.0
-11.6
48.0
0.6
32.0
11.8
40.375
36.6
40.0
-3.4
59.5
0.6
32.0
8.5
43.96
31.5
40.0
-8.5
55.7
0.7
32.0
7.1
49.97
38.1
40.0
-1.9
64.6
0.7
32.0
4.8
51.12
36.8
40.0
-3.2
63.4
0.7
32.0
4.7
60.0
29.7
40.0
-10.3
57.1
0.8
32.0
3.9
64.01
27.5
40.0
-12.5
53.8
0.8
32.0
4.9
79.15
29.4
40.0
-10.6
52.8
0.9
32.0
7.7
105.1
25.9
40.0
-14.1
51.1
1.0
32.0
5.7
Harmonic Emissions
IEC 61000-3-2
Not applicable. EUT AC mains has rated input currents exceeding 16 amps
per phase and is not directly connected to public power network.
Voltage Fluctuations/
Flicker emissions
Not applicable. EUT AC mains has rated input currents exceeding 16 Amps
per phase and is not directly connected to public power network.
The EQUIPMENT is suitable for use in all establishments other than domestic and those directly
connected to the public low-voltage power supply network that supplies buildings used for domestic
purposes.
Guidance and manufacturer’s declaration - electromagnetic emissions
The EQUIPMENT is intended for use in the electromagnetic environment specified below. The
customer or the user of the EQUIPMENT should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test level
Compliance level
Electromagnetic environment -
guidance