D000161-00 Rev. C
11
SECTION 2: SETUP AND INSTALLATION
The Anti-Gravity Treadmill will be installed by an AlterG qualified technician after delivery.
Please make sure that you inspect the Anti-Gravity Treadmill upon delivery for any damage that
may have occurred during transportation. Take pictures and immediately report any damage to
the shipping company and AlterG. When you sign for the shipment of your Anti-Gravity
Treadmill you are taking responsibility for any damage that may occur before installation.
Electrical Requirements
The recommended electrical power connection for the Anti-Gravity Treadmill is a dedicated 30
ampere, 220VAC @ 50/60 Hz circuit with ground. The plug supplied with the Anti-Gravity
Treadmill is designated by the NEMA configuration system as L6-30P. The corresponding
receptacle for the plug is a NEMA L6-30R.
International systems are shipped with a US plug on the power cord. An appropriate plug for
the country and facility in question should be wired as follows:
Blue Conductor: Neutral
Brown Conductor: Line
Green/Yellow Conductor: Ground
If Stride Smart is being installed, a 110V power outlet is required within ten feet of the front of the AlterG
Anti-Gravity Treadmill to power the display monitor.
Grounding Requirements
The Anti-Gravity Treadmill must be grounded electrically. If there is an electrical malfunction,
grounding provides a path of least resistance for electric current to reduce the risk of electric
shock. The Anti-Gravity Treadmill power cord includes a grounded plug. This plug must be
plugged into an appropriate receptacle (NEMA L6-30R) that is properly installed and grounded
in accordance with the current National Electrical Code as well as all local codes and
ordinances. If you are at all unsure of these requirements, contact AlterG or a qualified
electrician.
Location Requirements
The Anti-Gravity Treadmill measures 4ft (1.2m) x 8ft (2.4m) and weighs almost 1000 pounds. It
needs to be placed on a structurally sound surface. If it is used above ground level it is
advisable to place it near the corner of the room or where the floor will be strongest to ensure
maximum support during high-speed, high-impact use. The surface should be level to ensure
minimal flexing of the Anti-Gravity Treadmill frame. Do not place the Anti-Gravity Treadmill
directly on thick carpeting because it may interfere with air-valves located on the underside of
the machine. The Anti-Gravity Treadmill needs to be within 8 ft (2.4m) (from the front of the
machine) of the proper electrical outlet. Check with a qualified electrician or AlterG if you plan
on extending the cord in any way. Make sure you leave at least 24 inches on either side of the