48
Optical Fusion Splicer Instruction Manual
Bad quality of
splicing spot
1. Dust on fiber
2. Choose the wrong
fiber type or splice
mode
1. Re-prepare optical fiber
and re-splice
2. Choose the right fiber
type or splice mode
3. Do ARC Calibration, adjust
to proper ARC strength.
4. Do “Detect System
Parameter”
No-response
of keyboard
Abnormal system
running
Power off then re-start
fusion splicer
No light or
color
confusion for
screen
1. Abnormal system
running
2. Screen ribbon
cable loosen or
damage
Power off then re-start
fusion splicer, if the problem
is not solved, contact with
factory directly
The fibers
aren’t spliced
after running
splicing
ARC strength
Abnormal or
malfunction of
system running
Do ARC Calibration and re-
splice, if the problem is not
solved, contact with factory
directly
The fiber
exceed the
view field in
the alignment
process
1. Fiber is load out of
V-groove
2. Abnormal system
running
1. Re-load the fiber and
make sure it in the center of
V-groove
2. Main menu Splice Mode-
Splice Mode-Manual Enter,
user move left &right fiber
to the view field center, and
then do “Detect System
Parameter”