![Air Monitor MASS-tron II/CEM Скачать руководство пользователя страница 60](http://html1.mh-extra.com/html/air-monitor/mass-tron-ii-cem/mass-tron-ii-cem_installation-operation-and-maintenance-manual_2875334060.webp)
MASS-tron II/CEM
Air Monitor Corporation
116-004-94.P65 (8-31-10)
MASS-tron II/CEM
- IO&M Manual
SECTION 6 – CALIBRATION
55
6.3 – TRANSMITTER INPUT CALIBRATION (con't)
6.3.2
(con't)
Note: If span input is less than 80% or greater than 120% of Maximum Pressure input (see Factory Set-Up
Information Sheet), display will indicate:
Bad Input Span
-- Push ESCAPE --
If this occurs, check input pressure (as read on gauge), and readjust as necessary. Press
ESC
and repeat
Steps 9 through 10.
11. Press
ESC
.
12. Remove test equipment.
Содержание MASS-tron II/CEM