![Air Monitor MASS-tron II/CEM Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/air-monitor/mass-tron-ii-cem/mass-tron-ii-cem_installation-operation-and-maintenance-manual_2875334020.webp)
MASS-tron II/CEM
Air Monitor Corporation
116-004-94.P65 (8-31-10)
MASS-tron II/CEM
- IO&M Manual
15
SECTION 5 – OPERATION
5.5 – PUSHBUTTON DEFINITION
In addition to Configuration programming, pushbuttons can be used for certain functions when in the
Normal
operation
mode. The following list describes the pushbutton function when in the
Normal
operation mode and in the Operator
Setup (programming).
When in Normal Operation Mode.
ESC + ENT
Activates the Operator Setup menu.
ESC + UP
Displays a list of Manual Actions available on the top line of the display. Use
UP
or
DN
to scroll
through selections.
The following list describes all possible actions. Those available depend on installed options.
– Turn On AutoSpan Vlv
Output not held.
– Turn On AutoZero Vlv
Output not held.
– Perform Auto-Zero
Output held.
– Perform Auto-Calib.
Output held.
UP + DN
Displays software Version, W.O. #, S/N, and ID.
When in Operator Setup.
UP
or
DN
Use to scroll to the desired Main Menu item.
Use to scroll to the value or mode within a Parameter.
ENT
Enters user into specific selection sub-menu from Main Menu Selection.
Displays current setting of selected Parameter.
Enters the selected value or mode into memory.
ESC
Use to exit Main Menu selection to avoid scrolling to "Return to MAIN MENU".
The following pushbutton combinations can be used to more quickly set user selected values (i.e. display zero and
span, K-factor, Scaling Values, etc.).
UP* + ESC
Increase the second column digit.
DN* + ESC
Decrease the second column digit.
UP* + DN
Increase the third column digit.
X X X X X
DN* + UP
Decrease the third column digit.
UP* + ESC and DN
Increase the fourth column digit.
DN* + ESC and UP
Decrease the fourth column digit.
*Must be pressed and held before other button(s) are pressed.
2nd Column
3rd Column
4th Column
NOTE
Pushbuttons are momentary type and should be quickly pressed and released to
initiate desired change, unless otherwise instructed to press and hold.
Содержание MASS-tron II/CEM
Страница 2: ......