5.19 – TRANSDUCER SPAN SELECTION
Displays the transducer's natural span and operating span. The natural span represents the maximum process pres-
sure that the transducer can accept. The displayed natural span is for user/technician reference only and should not be
changed unless a new transducer of different natural span has been installed.
The operating span represents the calculated process pressure for a specific application, and can be adjusted 40% to
100% of the natural span.
SECTION 5 – OPERATION
46
1. While in Main Menu, use
UP
or
DN
to scroll to:
Transducer Span
Selection
2. Press
ENT
to enter Transducer Span Selection menu. Display will indicate:
Xdcr Natural Span
Selection
3. Press
ENT
and display indicates the natural span of the installed transducer (0.05, 0.10, 0.25, 0.50, 0.75, 1.00,
1.50, 2.00, 2.50, 3.00, 5.00, 10.00, 15.00, 25.00, 50.00, 100.00, 150.00, 200.00 or 250.00 IN w.c.).
4. Use
UP
or
DN
to change setting. Once desired setting is displayed, press
ENT
. New setting will be stored in
memory and display will return to Main Menu.
Caution: Only change setting if a transducer of different natural span has been installed.
Note: If user desires not to change the setting and return to Main Menu, press
ESC
. Unit will remain
programmed as it was originally.
- Analog Output Configuration
Transducer Span Selection
- Transmitter Input Calib.
- Xdcr Natural Span Selection
- Xdcr Operating Span Selection
- Return to MAIN MENU
Transducer Span Selection
Xdcr Natural Span
0.25 In. WC
Xdcr Operating Span
0.2500 In. WC
Return to
MAIN MENU
0.05, 0.10, 0.25, 0.50,
0.75, 1.00, 1.50, 2.00,
2.50, 3.00, 5.00, 10.00,
15.00, 25.00, 50.00,
100.00, 150.00, 200.00,
or 250.00
Adjustable from 40% to
100% of selected Natural
Span.
Default selection shown.
Available selections:
Operator Setup Main Menu