Running Samples
90
NovoCyte Quanteon Flow Cytometer Operator's Guide
Figure 58.
Import Samples from FCS Files
3
Ensure all the selected FCS files are successfully imported to the software.
Click
OK
to continue.
Figure 59.
Dialog Box for Successful Importing FCS Files
4
Set the compensation parameters in the
New Auto Compensation
window
(i.e.
Area
or
Height
,
Median
or
Mean
). Ensure the correct channel parameter is
selected for each single stained sample. To select different single stained
sample file, click the file name, select the desired file from the pull-down menu.