System Maintenance
NovoCyte Quanteon Flow Cytometer Operator's Guide
127
Figure 105.
Decontamination Function in NovoExpress
2
NovoExpress software will prompt a dialog box to guide you through the
decontaminating procedure. Follow the instructions in each step and click
OK
to move to the next step until completion.
Figure 106.
NovoExpress Dialog Box to Guide the Decontamination Procedure
NOTE
Contact your local Agilent representative or distributor for any questions regarding
this procedure.
Prime Fluidic System before Long-Term Storage
If there is no plan to run the instrument for more than two weeks, a thorough
cleaning process is recommended to prevent salt crystallization in the fluidic
system.
1
When the instrument is in
Ready
status, replace all the instrument reagents
with Milli-Q or 0.2 µm filtered DI water.