
Troubleshooting
134
NovoCyte Quanteon Flow Cytometer Operator's Guide
NovoFlow container is empty Refill the NovoFlow container following the
procedures described in associated instrument
operator's guide.
NovoFlow container is not
correctly connected to the
instrument and the NovoFlow
solution does not flow through
the tubing
Reconnect the NovoFlow quick coupling
connectors to ensure the NovoFlow container is
correctly connected to the instrument.
Execute
Prime
function from the NovoExpress
Software.
Sample Injection Probe is
clogged
Run NovoClean as sample to remove the things
clogged in SIP.
Flow cell is clogged
Execute the
Unclog
function from the
NovoExpress Software.
Sample tubing is leaking
Clean the leaked fluid with wipes and properly
reconnect the sample tubing.
Air bubbles exist in the
Sampling Pump
Execute
Debubble
function from NovoExpress
Software.
High CV
Air bubbles exist in the flow
cell
Execute the
Debubble
function from
NovoExpress Software.
Flow cell is clogged
Execute the
Unclog
function from NovoExpress
software
Improper sample preparation Re-prepare the sample properly.
The sample flow rate is set
too high
Lower the sample flow rate setting in the
NovoExpress software.
Excessive debris in
display
Sheath in-line filter is dirty
Replace the sheath in-line filter following the
procedures described in associated instrument
operator's guide.
The threshold is set too low
Increase the threshold setting and re-collect the
sample.
Fluidics system tubing is dirty
•
Execute
Rinse
function from NovoExpress
Software.
•
Execute
Cleaning
function from
NovoExpress Software.
•
Execute
Decontamination
function from
NovoExpress Software.
High event rate
The threshold is set too low
Increase the threshold setting and re-collect the
sample.
Sample concentration is too
high
Re-prepare the sample to appropriate
concentration or dilute the sample.
Sample flow rate is too high
Reduce the sample flow rate.
NovoExpress
displays
Instrument
Workstation is in hibernation
causing the interruption of
•
Restart NovoExpress Software.
•
Restart workstation.
Содержание NovoCyte Quanteon
Страница 10: ...Contents NovoCyte Quanteon Flow Cytometer Operator s Guide 10 This page is intentionally left blank...
Страница 28: ...Prologue 28 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 70: ...Instrument Operation 70 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 106: ...Running Samples 106 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 136: ...Troubleshooting 136 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 138: ...Version History 138 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 147: ......