Introduction
40
NovoCyte Quanteon Flow Cytometer Operator's Guide
Figure 13.
Hydrodynamic Focusing and Core Diameter under Different Sample Flow Rates
Having at most one cell or particle moving through a laser beam at a given time is
critical for flow cytometry. This is achieved by the hydrodynamic focusing process.
Proper operation of fluidic components is critical for particles to intercept the laser
beam properly. Always ensure that the fluidics system is free of air bubbles, cell
debris, and other contaminants. Choosing the appropriate sample flow rate for
your experiment is important for obtaining good quality data.
•
A higher sample flow rate is generally used for qualitative measurements such
as immunophenotyping. The data is less resolved but is acquired more
quickly.
•
A lower flow rate is generally used in applications where greater resolution and
quantitative measurements are critical, such as DNA content analysis.
There are three standard sample flow rate settings in Agilent flow cytometer,
which is Slow (e.g.
14 μL/min), Medium (
e.g.
35 μL/min) and Fast (
e.g.
66 μL/min),
corresponding to approximately 7.7 μm, 12.2 μm and 16.8 μm core diameter
respectively. The sample flow rate can also be continuously adjusted in
NovoExpress software from 5 μL/min to 120 μL/min, which corresponds to a core
diameter of 4.6 μm to 22.7 μm. Th
e sheath flow rate is fixed at 6.5 mL/min.
Содержание NovoCyte Quanteon
Страница 10: ...Contents NovoCyte Quanteon Flow Cytometer Operator s Guide 10 This page is intentionally left blank...
Страница 28: ...Prologue 28 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 70: ...Instrument Operation 70 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 106: ...Running Samples 106 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 136: ...Troubleshooting 136 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 138: ...Version History 138 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 147: ......