Instrument Operation
NovoCyte Quanteon Flow Cytometer Operator's Guide
65
6
Click
Restart
in the next window to restart the software to apply the new
optical configuration setting.
Figure 33.
Dialog Window for Restarting the NovoExpress
7
After NovoExpress is restarted, the following window will appear. Click
QC
Test
to continue.
Figure 34.
Dialog Window for Performing the QC Test
8
Properly prepare 1 mL QC particles sample as described in
NovoExpress
Software Guide
. Place the sample tube in the tube holder or NovoSampler Q.
Fill in the test information in the pop-up window. Click
Next
to continue.
Figure 35.
Fill in the Test Information