
Introduction
NovoCyte Quanteon Flow Cytometer Operator's Guide
37
principle. Although the signals differ in intensity, the percentage of the FITC
spillover into the PE detector remains constant.
Figure 11.
Spillover of Two FITC Signals of Different Intensities into PE Channels
The NovoExpress software provides flexible fluorescence compensation
functions. Users can modify the compensation matrix manually or use the
automatic compensation function. Fluorescence compensation can be conducted
during or after sample acquisition.
NOTE
For more information on how to conduct fluorescence compensation, please refer
to
Perform Fluorescence Compensation
NovoExpress Software
Guide
for details.
Fluidics System
Agilent flow cytometer fluidics system consists of sample injection probe (SIP),
SIP cleaning apparatus, driving pumps, flow cell, solenoid valves, and a series of
tubings and connectors.
Содержание NovoCyte Quanteon
Страница 10: ...Contents NovoCyte Quanteon Flow Cytometer Operator s Guide 10 This page is intentionally left blank...
Страница 28: ...Prologue 28 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 70: ...Instrument Operation 70 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 106: ...Running Samples 106 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 136: ...Troubleshooting 136 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 138: ...Version History 138 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 147: ......