Instrument Operation
60
NovoCyte Quanteon Flow Cytometer Operator's Guide
CAUTION
When adding reagents and emptying the waste, be sure to clean residual
liquid on the outside of the containers and place the container onto the
appropriate position in the NovoCyte Fluidics Station II.
WARNING
Biological samples may be biohazardous. Appropriate personal protective
equipment should be worn when handling the waste, which should be
disposed of in accordance with local safety regulations.
Empty Waste
If the waste container is more than 2/3 capacity, follow the instructions below to
empty the waste container.
NOTE
The procedures described below may vary if NovoCyte Fluidics Cart is used.
Please refer to
NovoCyte Fluidics Cart Operator's Guide
for more details.
1
Make sure the instrument is in
Ready
status or the power is turned off.
2
Disconnect the tubing from the waste container by gently pressing the metal
clip of the quick coupler on the cap.
3
Move the container from the NovoCyte Fluidics Station II to a sink. Unscrew
the container cap and pour the waste to a tank for disposing of biohazardous
materials.
4
Add 300 mL (i.e. 10% of total volume) bleach to the waste container.
5
Screw the container cap back onto the container.
6
Reconnect the quick coupler on the cap and place the container onto the
NovoCyte Fluidics Station II. Be careful to prevent any twisting or kinking of the
tubing.
CAUTION
Please make sure the instrument is in Ready status or powered off before
emptying the waste.
WARNING
Biological samples may be biohazardous. Appropriate personal protective
equipment should be worn when handling the waste, which should be
disposed of in accordance with local safety regulations.
Содержание NovoCyte Quanteon
Страница 10: ...Contents NovoCyte Quanteon Flow Cytometer Operator s Guide 10 This page is intentionally left blank...
Страница 28: ...Prologue 28 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 70: ...Instrument Operation 70 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 106: ...Running Samples 106 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 136: ...Troubleshooting 136 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 138: ...Version History 138 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 147: ......