System Maintenance
NovoCyte Quanteon Flow Cytometer Operator's Guide
117
Once the
Priming
procedure is completed, a dialog box below will be prompted for
conducting the QC Test. Prepare the QC particle sample as described in
NovoExpress Software Guide
and place it in the tube holder. Click
QC Test
to
initiate the QC Test. The QC Test will automatically start. Follow the steps and
complete the QC Test. A QC Test must be conducted each time when fluidic
system consumables are replaced in order to check the performance of the
system.
Figure 96.
Software Message Instructing a QC Test after Completion of Fluidic System Consumables
Replacement
The following sections describe the detailed procedure for replacing fluidic system
consumables. It is highly recommended to replace all the fluidic system
consumables at the same time to ensure the optimal system performance.
However, each component can also be replaced individually if necessary. Click
Instrument
>
Replace Fluidic System Consumables
in the NovoExpress software
to initiate the process. Always conduct the
Priming
process after the fluidic
system consumables have been replaced as prompted by the NovoExpress
software or manually conduct the
Priming
process by clicking
Instrument
>
Fluidics Maintenance
>
Priming
.
Replace the NovoFlow Sheath Fluid In-Line Filter
1
Make sure the instrument is in
Ready
or shut down status.
2
Open the front panel of the instrument. Locate the sheath fluid in-line filter as
shown below.
Содержание NovoCyte Quanteon
Страница 10: ...Contents NovoCyte Quanteon Flow Cytometer Operator s Guide 10 This page is intentionally left blank...
Страница 28: ...Prologue 28 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 70: ...Instrument Operation 70 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 106: ...Running Samples 106 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 136: ...Troubleshooting 136 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 138: ...Version History 138 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 147: ......