NovoCyte Quanteon Technical Specifications
144
NovoCyte Quanteon Flow Cytometer Operator's Guide
Table 15 NovoCyte Quanteon Technical Specifications
1
Sample Flow Rate
5-120 µL/min, continuously adjustable
Average Sheath Flow Rate
6.5 mL/min
Sample Aspiration Volume
5 µL -5 mL
Sample Recovery
Available
Compatibility to Autosampler
No fluidic tubing disassembly or re-connection
required.
Fluid Level Sensing
Active sensing using weight sensors with
automated warnings when any fluid level is out
of specified range.
Fluid Container Capacity
3 L sheath, 3 L waste, 500 mL cleaning, 500 mL
rinse. Optional large container for sheath (15 L)
and waste (15 L) with Fluidics Cart.
Carryover
< 0.1%
Sample Injection Probe (SIP)
Rinse
Automated flying collar wash of inner and outer
SIP surface after each sampling.
Fluidics System Monitoring
In-line pressure sensor monitors the pressure in
real time. Automated system recovery when
pressure is out of range due to clogging.
Fluidics Maintenance
Automated startup and shutdown with fluidic
system cleaning. Automated user executable
cleaning, debubble, rinse, unclog, priming and
decontamination. Automated execution of user
definable maintenance sequence.
Data
Acquisition
Software
NovoExpress
Parameters
Height and Area for FSC, SSC and all
Fluorescent Channels, Width off FSC and Time.
Dynamic Range
24 bits; 7.2 decades logarithmic scale
Fluorescence Photodetector
Gain Control
0-1000 User adjustable, optimized default gain
setting for each individual channel.
Compensation
Full inter-beam matrix, during or post
acquisition.
File Export
FCS 3.0, FCS 3.1; CSV; Batch PDF report
Data Report
Automatic report. Customizable report. Batch
PDF report.
Usage Monitor
Comprehensive Transaction Log and System
Log.
Содержание NovoCyte Quanteon
Страница 10: ...Contents NovoCyte Quanteon Flow Cytometer Operator s Guide 10 This page is intentionally left blank...
Страница 28: ...Prologue 28 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 70: ...Instrument Operation 70 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 106: ...Running Samples 106 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 136: ...Troubleshooting 136 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 138: ...Version History 138 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 147: ......