System Maintenance
NovoCyte Quanteon Flow Cytometer Operator's Guide
109
Figure 82.
Software Message Instructing to Remove Sample Tube
WARNING
Sample injection probe will move downward after clicking the OK button.
Do not put your fingers underneath the sample injection probe!
5
Prepare a cotton swab or a damp soft cloth with NovoClean solution (1X) and
gently wipe the outer surface of the sample injection probe.
Figure 83.
Wipe the Outer Surface of the Sample Injection Probe