50
Chapter 2
Performance Test
E5071C Performance Test Record
E5071C Performance Test Record
Agilent Technologies E5071C RF Network Analyzer
Serial Number:
Option:
Temperature:
°
C
Test Date:
Humidity:
% R.H.
Tested by:
AUX Input Test
1. Multimeter : 3458A
AUX1 Offset error
AUX Power
Range[V]
Test limit [mV]
Test result [mV]
Measurement
Uncertainty [mV]
1
±
1
±
0.01
10
±
10
±
0.05
AUX1 Proportional error
AUX Input
Range[V]
Test limit [mV]
Test result [mV]
Measurement
uncertainty [mV]
1
±
11
±
0.52
- 1
±
11
±
2.3
10
±
110
±
1.3
- 10
±
110
±
1.2
AUX2 Offset error
AUX Power
Range[V]
Test limit [mV]
Test result [mV]
Measurement
Uncertainty [mV]
1
±
1
±
0.01
10
±
10
±
0.05
AUX2 Proportional error
AUX Input
Range[V]
Test limit [mV]
Test result [mV]
Measurement
uncertainty [mV]
1
±
11
±
0.52
- 1
±
11
±
2.3
Содержание E5071C
Страница 14: ...14 Contents ...
Страница 168: ...168 Chapter2 Performance Test E5071C Performance Test Record ...
Страница 212: ...212 Chapter4 Troubleshooting To configure the CPU Mother Board Assembly and BIOS ...
Страница 262: ...262 Chapter5 Replaceable Parts Power Cables and Plug Configurations ...
Страница 326: ...326 Chapter6 Replacement Procedure Probe Power and Power Switch Board Replacement ...
Страница 332: ...332 Chapter7 Post Repair Procedures Post Repair Procedures ...
Страница 360: ...360 AppendixB System Recovery Calibration of the Touch Screen ...
Страница 364: ...364 AppendixC Firmware Update Firmware Update ...
Страница 365: ...365 D Power Requirement D Power Requirement ...
Страница 367: ...Appendix D 367 Power Requirement Preparation for Power Supply D Power Requirement Figure D 1 Power cable options ...