![Agilent Technologies E5071C Скачать руководство пользователя страница 33](http://html.mh-extra.com/html/agilent-technologies/e5071c/e5071c_service-manual_2868073033.webp)
Chapter 2
33
Performance Test
4. RF OUTPUT LEVEL LINEARITY TEST
2. Pe
rfor
ma
nce T
est
4. RF OUTPUT LEVEL LINEARITY TEST
Description
This test checks the level accuracy of the E5071C test port output signal across the
specified level range. The RF output level is measured for power level settings of -20 dBm
to 10 dBm in 2.5dB step increments at 9 kHz, 100 kHz, 50 MHz, 3 GHz, 4.5 GHz, 5 GHz,
and -20 dBm to 9 dBm in 2.5 dB step increments at 6 GHz, and -20 dBm to 8 dBm in 2.5
dB step increments at 7 GHz, and -20 dBm to 7 dBm in 2.5 dB step increments at 8.5 GHz.
Test equipment
Power meter
Agilent E4419A/B
Power sensor
Agilent E9304A Opt. H18 or Agilent E9304A Opt.
H19
Содержание E5071C
Страница 14: ...14 Contents ...
Страница 168: ...168 Chapter2 Performance Test E5071C Performance Test Record ...
Страница 212: ...212 Chapter4 Troubleshooting To configure the CPU Mother Board Assembly and BIOS ...
Страница 262: ...262 Chapter5 Replaceable Parts Power Cables and Plug Configurations ...
Страница 326: ...326 Chapter6 Replacement Procedure Probe Power and Power Switch Board Replacement ...
Страница 332: ...332 Chapter7 Post Repair Procedures Post Repair Procedures ...
Страница 360: ...360 AppendixB System Recovery Calibration of the Touch Screen ...
Страница 364: ...364 AppendixC Firmware Update Firmware Update ...
Страница 365: ...365 D Power Requirement D Power Requirement ...
Страница 367: ...Appendix D 367 Power Requirement Preparation for Power Supply D Power Requirement Figure D 1 Power cable options ...