26
Chapter 1
General Information
Power Meter Accuracy Test
TEST RECORD FOR POWER METER ACCURACY TEST
Power Meter Tested
Model Number:
Test Date:
Serial Number:
Tested by:
Test Equipment Used
Range Calibrator
: Model No.:
Serial No.:
Digital Voltmeter
: Model No.:
Serial No.:
Test Results
Range
Calibrator:
Setting
Channel A
Channel B
DVM Reading
(Vdc)
Power Meter
Reading (dBm)
DVM Reading
(Vdc)
Power Meter
Reading (dBm)
1 mW
A=
0.00
±
0.001
A=
0.00
±
0.001
300
μ
W
B =
C =
B =
C=
100
μ
W
D =
E=
D =
E =
Pass/Fail Calculation
300
μ
W
R=B/A =
S=
10
(C/10)
=
% ERROR=((R-S)/R)x100 =
Limits:
±
0.13%
Pass __ Fail __
R=B/A =
S=
10
(C/10)
=
% ERROR=((R-S)/R)x100 =
Limits:
±
0.13%
Pass __ Fail __
100
μ
W
T=D/A =
U=
10
(E/10)
=
% ERROR=((T-U)/T)x100 =
Limits:
±
0.10%
Pass __ Fail __
T=D/A =
U=
10
(E/10)
=
% ERROR=((T-U)/T)x100 =
Limits:
±
0.10%
Pass __ Fail __
Содержание E5071C
Страница 14: ...14 Contents ...
Страница 168: ...168 Chapter2 Performance Test E5071C Performance Test Record ...
Страница 212: ...212 Chapter4 Troubleshooting To configure the CPU Mother Board Assembly and BIOS ...
Страница 262: ...262 Chapter5 Replaceable Parts Power Cables and Plug Configurations ...
Страница 326: ...326 Chapter6 Replacement Procedure Probe Power and Power Switch Board Replacement ...
Страница 332: ...332 Chapter7 Post Repair Procedures Post Repair Procedures ...
Страница 360: ...360 AppendixB System Recovery Calibration of the Touch Screen ...
Страница 364: ...364 AppendixC Firmware Update Firmware Update ...
Страница 365: ...365 D Power Requirement D Power Requirement ...
Страница 367: ...Appendix D 367 Power Requirement Preparation for Power Supply D Power Requirement Figure D 1 Power cable options ...