4
Operating in EI Mode
86
5975/5977 Series MSD Operating Manual
LCP Status Messages
The following messages may be displayed on the LCP to inform you of the
status of the MSD system. If the LCP is currently in Menu mode, cycle through
the menus to return to Status mode. No messages will be displayed if data
acquisition software is not controlling the MSD.
ChemStation Loading <timestamp>
The Agilent OpenLAB CDS software is starting up.
Executing <type>tune
A tuning procedure is in progress (type = QuickTune or Autotune).
Instrument Available <timestamp>
The Agilent OpenLAB CDS software is not running.
Loading Method <method name>
Method parameters are being sent to the MSD.
Loading MSD Firmware
The MSD’s firmware is being initialized.
The following messages alternately appear on the LCP if the MSD does
NOT
complete its bootup sequence properly:
Server not Found
Check LAN Connection
Seeking Server
Bootp Query xxx
These messages indicate that the MSD has not received its unique IP address
from the Windows Service. If the messages persist after you have logged onto
your account in the Agilent OpenLAB CDS program, consult the Troubleshooting
section of the Software Installation manual.
Loading OS
The operating system of the instrument controller is being initialized.