5
General Maintenance
158
5975/5977 Series MSD Operating Manual
5
Secure the new filaments with the gold plated screw and washer. (See
6
After installing the filament, verify that it is not grounded to the source body.
7
Install the EI source. (See
“To Install the EI XTR, SS, or Inert Source”
8
Close the analyzer chamber. (See
“To Close the Analyzer Chamber”
9
Pumpdown the MSD. (See
10
Autotune the MSD. (See
11
In the Manual Tune dialog, the
Filament
parameter allows you to enter
1
or
2
for the filament number. Whichever number was present during the previous
autotune enter the other filament number.
12
Autotune the MSD again.
13
Enter the filament number that gave the best results.
If you decide to use the first filament number, run Autotune again to make
sure the tune parameters are compatible with the filament.
14
Select
Save Tune Parameters
from the
File
menu.