5
General Maintenance
5975/5977 Series MSD Operating Manual
129
4
Abrasively clean the surfaces that contact the sample or ion beam.
Use an abrasive slurry of alumina powder and reagent-grade methanol on a
cotton swab. Use enough force to remove all discolorations. Polishing the
parts is not necessary; small scratches will not harm performance. Also,
abrasively clean the discolorations where electrons from the filaments
enter the source body.
5
Rinse away all abrasive residue with reagent-grade methanol.
Ensure
all
abrasive residue is rinsed away
before
ultrasonic cleaning. If the
methanol becomes cloudy or contains visible particles, rinse again.
6
Separate the parts that were abrasively cleaned from the parts that were not
abrasively cleaned.
7
Ultrasonically clean the parts (each group separately) for 15 minutes in each
of the following solvents:
•
Methylene chloride (reagent-grade)
•
Acetone (reagent-grade)
•
Methanol (reagent-grade)
8
Place the parts in a clean beaker.
Loosely
cover the beaker with clean
aluminum foil (dull side down).
9
Dry the cleaned parts in an oven at 100°C for 5–6 minutes.
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
WARNING
All of these solvents are hazardous. Work in a fume hood, and take all
appropriate precautions.