2
Installing 7890/7820/6890/6850 GC Columns
48
5975/5977 Series MSD Operating Manual
Calculation for average linear velocity
Average linear velocity (cm/s)
=
where:
L
= Length of the column in meters
t
= Retention time in seconds
Calculation for volumetric flow rate
Volumetric flow rate (mL/min)
=
where:
D
= Internal column diameter in millimeters
L
= Column length in meters
t
= Retention time in minutes
100
L
t
---------------
0.785
D
2
L
t
---------------------------