4
Operating in EI Mode
100
5975/5977 Series MSD Operating Manual
View MSD Temperatures and Vacuum
Pressure monitoring requires an optional Micro-Ion vacuum gauge
(G3397A for
the 5975 and G3397B for the 5977 series).
Select
Instrument Status
in the windows area and click
MSD
to display the MSD’s
actuals values including the source and quad temperatures and quad manifold
vacuum pressure.
WARNING
If you are using hydrogen as a carrier gas, do not turn on the Micro-Ion
vacuum gauge if there is any possibility that hydrogen has accumulated in the
analyzer chamber. Read
on page 19 before operating the
MSD with hydrogen carrier gas.
Figure 26.
The MSD Instrument Status screen