1
Introduction
5975/5977 Series MSD Operating Manual
27
To Replace the Primary Fuses
Materials needed
•
Fuse, T12.5A, 250 V (2110-1398) – 2 required
•
Screwdriver, flat-blade (8730-0002)
The most likely cause of primary fuse failure is a problem with the foreline pump.
If the primary fuses in your MSD fail, check the foreline pump.
Procedure
1
Vent the MSD, and unplug the power cord from the electrical outlet.
If one of the primary fuses has failed, the MSD will already be off, but for
safety, switch off the MSD and unplug the power cord. It is not necessary to
allow air into the analyzer chamber.
2
Turn one of the fuse holders counterclockwise until it pops out. The fuse
holders are spring loaded. (See
3
Remove the old fuse from the fuse holder.
4
Install a new fuse in the fuse holder.
5
Reinstall the fuse holder.
WARNING
Never replace the primary fuses while the MSD is connected to a power
source.
WARNING
If you are using hydrogen as a carrier gas, the hydrogen gas flow must be off
before turning off the MSD power. If the foreline pump is off, hydrogen will
accumulate in the MSD and an explosion may occur. Read
on page 19 before operating the MSD with hydrogen gas.