5
General Maintenance
5975/5977 Series MSD Operating Manual
125
11
Remove the five lenses from the lens insulator/holder.
CAUTION
Use care when removing the lenses from the lens insulator casing. Putting
excessive stress on this casing can break or crack it. If this happens, do not
attempt to operate with a defective lens insulator, it must be replaced.
CAUTION
Use care when removing the filament from the filament block. Putting
excessive stress on this can break or crack it. If this happens, do not attempt to
operate with a defective filament, it must be replaced.
Figure 37.
Exploded parts view of the EI HES
1
2
8
16
17
15
9
4
3
11
10
6
5
7
13
12
14