3
Intuvo 9000 GC Column Maintenance
5975/5977 Series MSD Operating Manual
71
4
Gently
check the alignment of the analyzer and the interface.
When the analyzer is aligned correctly, the analyzer can be closed all the
way with no resistance except the spring tension from the interface tip seal.
5
You can align the analyzer and interface by wiggling the side plate on its
hinge. If the analyzer still will not close, contact your Agilent Technologies
service representative.
CAUTION
Forcing the analyzer closed if these parts are misaligned will damage the seal
or the interface or the ion source, or keep the sideplate from sealing.