5
General Maintenance
5975/5977 Series MSD Operating Manual
141
To Disassemble the EI SS or EI Inert Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Hex ball driver, 2.0 mm (8710-1804)
•
Wrench, open-end, 10 mm (8710-2353)
Procedure
Refer to the exploded parts view
and the EI SS and EI Inert parts list
on page 142 while using this procedure.
1
Remove the EI source. (See
Remove the EI XTR, SS, or Inert Source”
2
Remove the two gold plated screws from the filaments and remove the
filaments from the source.
3
Loosen the two gold plated screws from the source heater block assembly,
and separate the repeller assembly from the source body. The repeller
assembly includes the source heater block assembly, repeller, and related
parts.
4
Remove the repeller nut and washers then remove the repeller from the
source heater block assembly.
5
Remove the repeller insulators and the repeller block insert from the source
heater block assembly.
6
Remove the gold plated setscrew from the side of the source body.
7
Push on the drawout plate to remove the entrance lens, ion focus lens,
drawout cylinder, and drawout plate from the other end of the source body.
8
Unscrew the interface socket. A 10-mm open-end wrench fits the flats on the
interface socket.
9
Remove the entrance lens and ion focus lens from the lens insulator.
Содержание 5975 Series
Страница 1: ...Agilent 5975 5977 MSD for OpenLAB CDS Operating Manual ...
Страница 74: ...3 Intuvo 9000 GC Column Maintenance 74 5975 5977 Series MSD Operating Manual ...
Страница 163: ......
Страница 164: ...www agilent com Agilent Technologies Inc 2019 First Edition January 2019 G7077 90031 G7077 90031 ...