2
Installing 7890/7820/6890/6850 GC Columns
34
5975/5977 Series MSD Operating Manual
7
Position the column so it extends 4 to 6 mm past the end of the ferrule. (See
8
Slide the septum up to the bottom of the nut to fix the correct column
insertion length.
9
Insert the column in the inlet.
10
Slide the nut up the column to the inlet base, and finger-tighten the nut.
11
Adjust the column position so the septum is even with the bottom of the
column nut.
12
Tighten the column nut an additional 1/4 to 1/2 turn. The column should not
slide with a gentle tug.
13
Start carrier gas flow.
14
Verify flow by submerging the free end of the column in isopropanol. Look for
bubbles.
See also
For more information about installing a capillary column, refer to
Optimizing
Splitless Injections on Your GC for High Performance MS Analysis
, Agilent
Technologies publication number 5988-9944EN.
Figure 4.
Installing a capillary column for a split/splitless inlet
Insulation cup
Reducing nut
Capillary column
Ferrule (inside nut)
Inlet column nut
Septum
4 to 6 mm