4349B
DISPlay
Subsystem
DISPlay
Subsystem
The
DISPlay
subsystem
controls
the
selection
of
the
channel
displayed
and
the
display
mode
.
COMMAND
P
ARAMETER
DISPlay
:ENABle
<Boolean>
:WINDowf 1j 2j 3j4g
[:ST
A
T
e]
<Boolean>
:TEXT1
:DIGit
f 3j 4j 5g
:P
A
GE
f 1j 2g
:TEXT2
:P
A
GE
f 1j 2j 3g
Note
The
state
of
each
channel's
display
mode
,
display
digits
and
display
mode
of
the
Measurement
Settings
are
linked.
:DISPlay:ENABle
f
ON
j
OFF
j
1
j
0
g
Controls
or
queries
whether
the
display
is
ON
or
OFF
.
ON ,
1
set
display
visible
.
OFF ,
0
set
display
blank.
Query
response
is
{1|0} .
:DISPlay:WINDow f1j 2j 3
j 4g[:ST
A
T
e]
f
ON
j
OFF
j
1
j
0
g
Selects
the
channel
to
be
displayed
and
controls
whether
or
not
the
display
is
turned
ON,
or
queries
whether
or
not
the
selected
channel
is
set
to
ON.
WINDow{1|2|3|4}
selects
the
displayed
channel.
ON ,
1
displays
selected
channel.
OFF ,
0
sets
the
display
blank.
Query
response
is
{1|0} .
:DISPlay:WINDow f1j 2j 3
j 4g:TEXT1:P
A
GE
f
1
j
2
g
Selects
the
displayed
channel
and
display
mode
.
WINDow{1|2|3|4}
selects
the
displayed
channel.
1
displays
the
measurement
results
of
selected
channel.
2
displays
the
comparison
results
of
all
channels
.
Query
response
is
{1|2} .
GPIB
Reference
5-15
Содержание 4349B
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 20: ......
Страница 29: ...C Overload No Contact Operations Messages Instrument Errors Messages 2 GPIB Errors Messages 3 Index Contents 9 ...
Страница 42: ...4349B Figure 1 1 Power Cable Supplied 1 10 Getting Started ...
Страница 66: ......
Страница 88: ......
Страница 104: ......
Страница 146: ......
Страница 160: ......
Страница 173: ...4349B Figure 8 5 Current Measurement Additional Error Measurement Time 400 msec General Information 8 11 ...
Страница 176: ...4349B Figure 8 9 10 Additional Error Measurement Time 400 msec 8 14 General Information ...
Страница 210: ...4349B Figure B 1 Handler Interface Comparison Output Signals Diagram B 2 Handler Interface Installation ...
Страница 212: ...4349B Figure B 3 Handler Interface Input Signal Diagram B 4 Handler Interface Installation ...
Страница 216: ...4349B Figure B 4 A1 Main Board B 8 Handler Interface Installation ...
Страница 228: ......