4349B
ABORt
Command
ABORt
Command
:ABORt
The
ABORt
command
resets
the
trigger
system
and
places
all
trigger
sequences
in
the
IDLE
state
.
Any
actions
related
the
trigger
system
that
are
in
progress
,
such
as
acquiring
a
measurement,
are
aborted
immediately
.
The
execution
of
an
ABORt
command
will
set
any
the
pending
operation
ags
to
false
,
for
example
ags
that
were
set
by
the
initiation
of
the
trigger
system.
See
\Trigger
System
".
Unlike
*RST ,
ABORt
does
not
alter
the
settings
programmed
by
other
commands
.
(No
query)
Note
After
FETCh?
query
,
*TRG
command,
or
the
B
ASIC
command
TRIGGER ,
the
ABORt
command
causes
the
GPIB
bus
to
hang
up
.
T
o
avoid
this
,
before
ABORt
command
execution,
clear
the
gpib
bus
with
the
B
ASIC
command
CLEAR (address).
GPIB
Reference
5-9
Содержание 4349B
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 20: ......
Страница 29: ...C Overload No Contact Operations Messages Instrument Errors Messages 2 GPIB Errors Messages 3 Index Contents 9 ...
Страница 42: ...4349B Figure 1 1 Power Cable Supplied 1 10 Getting Started ...
Страница 66: ......
Страница 88: ......
Страница 104: ......
Страница 146: ......
Страница 160: ......
Страница 173: ...4349B Figure 8 5 Current Measurement Additional Error Measurement Time 400 msec General Information 8 11 ...
Страница 176: ...4349B Figure 8 9 10 Additional Error Measurement Time 400 msec 8 14 General Information ...
Страница 210: ...4349B Figure B 1 Handler Interface Comparison Output Signals Diagram B 2 Handler Interface Installation ...
Страница 212: ...4349B Figure B 3 Handler Interface Input Signal Diagram B 4 Handler Interface Installation ...
Страница 216: ...4349B Figure B 4 A1 Main Board B 8 Handler Interface Installation ...
Страница 228: ......