37
Chapter 3: Testing Performance
To test the single-clock, single-edge, state acquisition
To test the single-clock, single-edge, state acquisition
Testing the single-clock, single-edge, state acquisition verifies the performance of
the following specifications:
•
Minimum master-to-master clock time.
•
Maximum state acquisition speed.
•
Setup/Hold time for single-clock, single-edge, state acquisition.
This test checks two combinations of data channels using a single-edge clock at
two selected setup/hold times.
Equipment Required
Set up the equipment
If you have not already done so, do the following procedures:
“To set up the test equipment and the logic analyzer” on page 23.
“To set up the logic analyzer for the state mode tests” on page 33.
Connect and configure the logic analyzer
1
Using the 6-by-2 test connectors, connect the first combination of logic
analyzer clock and data channels listed in one of the following tables to the
pulse generator.
If you are testing a 1680/81/90/91A,AD, you will repeat this test for the second
combination.
Equipment
Critical Specifications
Recommended Model/Part
Pulse Generator
200 MHz 2.5 ns pulse width,
<
600 ps rise time 8133A option 003
Digitizing Oscilloscope
≥
6 GHz bandwidth,
<
58 ps rise time
54750A w/ 54751A
Adapter
SMA(m)-BNC(f)
1250-1200
SMA Coax Cable (Qty 3)
18 GHz bandwidth
8120-4948
Coupler
BNC(m)(m)
1250-0216
BNC Test Connector,
6x2 (Qty 4)
Содержание 1680 series
Страница 13: ...13 Chapter 1 General Information Dimensions 1680A AD Series 1690A AD Series...
Страница 15: ...15 2 Preparing for Use This chapter gives you instructions for preparing the logic analyzer for use...
Страница 74: ...74 Chapter 3 Testing Performance Performance Test Record...
Страница 77: ...77 5 Troubleshooting This chapter helps you troubleshoot the logic analyzer to find defective assemblies...
Страница 142: ...142 Chapter 6 Replacing Assemblies 1690A AD series disassembly assembly...
Страница 146: ...146 Chapter 7 Replaceable Parts Exploded View Exploded view of the Agilent 1680A AD series logic analyzer...
Страница 153: ...153 Chapter 7 Replaceable Parts Exploded View Exploded view of the Agilent 1690A AD series logic analyzer...
Страница 159: ...159 8 Theory of Operation This chapter tells the theory of operation for the logic analyzer and describes the self tests...
Страница 172: ...172 Chapter 8 Theory of Operation Self Tests Descriptions...
Страница 174: ......