111
Chapter 5: Troubleshooting
General Troubleshooting
analyzer Trigger In BNC.
c
Using a BNC-banana cable, connect the voltmeter to the logic analyzer
Trigger Out BNC.
The voltmeter will display a voltage approximately 3 Vdc.
4
Configure the external trigger:
a
Select the
Trigger Setup icon.
b
In the Advanced Trigger dialog, for Trigger Sequence Step 1, select
“Arm in from” instead of “Anything” as the event to trigger on.
c
Then, select “External trigger” as the source of the arming signal.
d
Click OK to close the Advanced Trigger dialog.
5
Verify the external trigger:
a
Select the
Run icon.
The logic analyzer will report "Waiting in Trigger Step 1...". The voltmeter will
display approximately 0 Vdc.
b
On the DC source, enter a voltage setting of 3.000 V.
The voltmeter will display approximately 3 Vdc. The logic analyzer will trigger
and display a waveform similar to the following:
c
On the DC source, re-enter the voltage setting of 0.000 V.
6
Disconnect the test equipment from the logic analyzer.
Содержание 1680 series
Страница 13: ...13 Chapter 1 General Information Dimensions 1680A AD Series 1690A AD Series...
Страница 15: ...15 2 Preparing for Use This chapter gives you instructions for preparing the logic analyzer for use...
Страница 74: ...74 Chapter 3 Testing Performance Performance Test Record...
Страница 77: ...77 5 Troubleshooting This chapter helps you troubleshoot the logic analyzer to find defective assemblies...
Страница 142: ...142 Chapter 6 Replacing Assemblies 1690A AD series disassembly assembly...
Страница 146: ...146 Chapter 7 Replaceable Parts Exploded View Exploded view of the Agilent 1680A AD series logic analyzer...
Страница 153: ...153 Chapter 7 Replaceable Parts Exploded View Exploded view of the Agilent 1690A AD series logic analyzer...
Страница 159: ...159 8 Theory of Operation This chapter tells the theory of operation for the logic analyzer and describes the self tests...
Страница 172: ...172 Chapter 8 Theory of Operation Self Tests Descriptions...
Страница 174: ......