115
Chapter 6: Replacing Assemblies
1680A,AD-series disassembly/assembly
7
Reverse this procedure to install the chassis into the sleeve.
When reassembling, check the following:
all assemblies are properly installed before installing the chassis into the sleeve.
ensure all exposed cables are dressed properly so the sleeve does not cause any
damage to the cables.
To remove the acquisition board
1
Do the procedure "To remove the chassis from the sleeve".
2
Turn the chassis upside down.
3
Disconnect the IEEE 1394 cable from the acquisition board.
4
Using a Torx T-10 screwdriver, remove five screws that secure the probe
shroud to the rear panel of the logic analyzer.
5
Using a Torx T-10 screwdriver, remove one screw that secures the
acquisition board to the chassis (center of the acquisition board).
Содержание 1680 series
Страница 13: ...13 Chapter 1 General Information Dimensions 1680A AD Series 1690A AD Series...
Страница 15: ...15 2 Preparing for Use This chapter gives you instructions for preparing the logic analyzer for use...
Страница 74: ...74 Chapter 3 Testing Performance Performance Test Record...
Страница 77: ...77 5 Troubleshooting This chapter helps you troubleshoot the logic analyzer to find defective assemblies...
Страница 142: ...142 Chapter 6 Replacing Assemblies 1690A AD series disassembly assembly...
Страница 146: ...146 Chapter 7 Replaceable Parts Exploded View Exploded view of the Agilent 1680A AD series logic analyzer...
Страница 153: ...153 Chapter 7 Replaceable Parts Exploded View Exploded view of the Agilent 1690A AD series logic analyzer...
Страница 159: ...159 8 Theory of Operation This chapter tells the theory of operation for the logic analyzer and describes the self tests...
Страница 172: ...172 Chapter 8 Theory of Operation Self Tests Descriptions...
Страница 174: ......