128
Chapter 6: Replacing Assemblies
1680A,AD-series disassembly/assembly
To disassemble the front panel assembly
Remove the front panel circuit board.
1
Remove each of the knobs from the front panel.
2
Using a Torx T-10 screwdriver, remove five screws that secure the front
panel circuit board to the front panel.
3
Lift the front panel circuit board out of the front panel.
4
Lift the elastomeric keypad out of the front panel.
Remove the LCD display.
1
Using a Torx T-10 screwdriver, remove four screws, one in each corner,
that secures the LCD display to the front panel.
2
Lift the LCD display out of the front panel.
3
Lift the glass lens out of the front panel assembly.
NOTE:
When installing, blow any dust off of the LCD and lens.
Remove the on/off switch.
1
Using a Torx T-10 screwdriver, remove two screws that secure the on-off
switch circuit board to the front panel.
2
Lift the on/off switch circuit board out of the front panel.
3
Lift the on/off switch out of the front panel.
Содержание 1680 series
Страница 13: ...13 Chapter 1 General Information Dimensions 1680A AD Series 1690A AD Series...
Страница 15: ...15 2 Preparing for Use This chapter gives you instructions for preparing the logic analyzer for use...
Страница 74: ...74 Chapter 3 Testing Performance Performance Test Record...
Страница 77: ...77 5 Troubleshooting This chapter helps you troubleshoot the logic analyzer to find defective assemblies...
Страница 142: ...142 Chapter 6 Replacing Assemblies 1690A AD series disassembly assembly...
Страница 146: ...146 Chapter 7 Replaceable Parts Exploded View Exploded view of the Agilent 1680A AD series logic analyzer...
Страница 153: ...153 Chapter 7 Replaceable Parts Exploded View Exploded view of the Agilent 1690A AD series logic analyzer...
Страница 159: ...159 8 Theory of Operation This chapter tells the theory of operation for the logic analyzer and describes the self tests...
Страница 172: ...172 Chapter 8 Theory of Operation Self Tests Descriptions...
Страница 174: ......